Verify certificate

The CE marking on a medical device

The CE marking shows that a product meets the requirements of EU legislation and may circulate freely in the European Union. On a device certified by MCA it is followed by our number, 3132.

3132
The marking keeps its proportions when scaled and is at least 5 mm high. The Notified Body number follows it.
Before the label goes to print
  • Proportions taken from the drawing in MDR Annex V
  • At least 5 mm high, unless the device is very small
  • 3132 placed right after the CE marking
  • On the device or its sterile packaging, the instructions for use and any sales packaging
Verify an MCA certificateCheck that a certificate behind a 3132 marking is valid
The marking itself

Rules for affixing the CE marking

  1. Proportions

    The marking keeps the proportions of the graduated drawing in the regulation when it is reduced or enlarged.

    MDR Annex V
  2. Minimum size

    The vertical dimension is at least 5 mm. This minimum may be waived for small-scale devices.

    MDR Annex V
  3. Where it goes

    Visibly, legibly and indelibly on the device or its sterile packaging. Where that is not possible or not suitable, on the packaging. It also appears in the instructions for use and on any sales packaging.

    MDR Article 20(3)
  4. Notified Body number

    Where a Notified Body took part in the conformity assessment, its identification number follows the CE marking. The number also appears in promotional material that says the device meets the CE marking requirements.

    MDR Article 20(5)
Reading a device label

Six things on a medical device label

Scroll through the list, or point at a part of the label, to see where each one sits.

  1. 1
    CE marking

    The manufacturer’s declaration that the device meets Regulation (EU) 2017/745.

    MDR Article 20
  2. 2
    Notified Body number

    Identifies the Notified Body responsible for the conformity assessment. For devices we certify, this is 3132.

    MDR Article 20(5)
  3. 3
    Unique Device Identifier

    The device identifier and production data, as a barcode and in plain text, so the device can be traced.

    MDR Article 27 · Annex VI
  4. 4
    Medical device symbol

    Shows that the product is a medical device, in every language.

    MDR Annex I §23.2 · ISO 15223-1
  5. 5
    Reference and serial number

    Catalogue number and serial number, shown with harmonised symbols, so a single device can be identified.

    ISO 15223-1
  6. 6
    Manufacturer

    Name and registered place of business of the legal manufacturer. Manufacturers outside the EU also show their authorised representative.